Pure and Ultrapure Water Sensor
This sensor is especially designed for pure and ultrapure water applications such as demineralization or desalination.
Sterilizable Sensor Design
The cell of the InPro7001-VP conductivity sensor is steam-sterilizable up to 131°C.
Meets Food & Beverage Industry Demands
This sensor meets the hygienic demands of food production. Its 316 stainless steel design and heat tolerance make it optimal for SIP processes.
Material No.: 52001997
Temperature Sensor | PT 1000 |
Cell Constant | 0.1 cm-1 |
Cable Connection | Vario Pin (IP 68) |
Cell Constant (nominal) | 0.1 cm–1 |
Measurement Technology | Conductive |
Process Connection | Pg 13.5 |
Sterilizable | Yes |
Operating Temperature (Fahrenheit) | 14...212 °F |
O-Ring Material | Viton® (FDA) |
Operating Temperature (Celsius) | -10...100 °C |
Mechanical Pressure Resistance (PSI) | 250 psi (77 °F), 100 psi (203 °F) |
Model Family | 2-electrode sensor |
Hazardous Area Approvals | Yes |
Measured Pressure Resistance | 0 – 100 psig @ 203 °F 0 – 246 psig @ 77 °F (7 bar @ 95 °C 17 bar @ 25 °C) |
Surface Roughness of Wetted Metal Parts | Polished N4 (Ra < 0.2 μm) (Ra < 8 μin) |
Electrode Material | SS 316L |
Primary Media/Application | Other Water Processes Pharmaceutical Waters |
Hygienic/Biocompatibility Approvals | Yes |
Mechanical Pressure Resistance (Bar) | 17 bar (25°C), 7 bar (95°C) |
Sensor Material | 316L SS |
Cell Constant (accuracy) | ± 1.0 % |
Measurement Range | 0.02-500 µS/cm |
Process Reliability and Easy Connection
The IP68 rated VarioPin connector system ensures excellent signal transmission and avoids problems associated with humidity.
Easy Process Installation
The InPro 7000-VP conductivity sensor can simply be screwed into a pipe-tee via the Pg 13.5 process connection and is immediately available to use.
Improved Process Safety
The sensor provides both improved process safety and enhanced support for validation and qualification procedures.