In-Line Gas Phase O₂ Measurement
The InPro 6800G/12/120 is designed to withstand harsh process conditions to provide precise, in-line, gas-phase oxygen measurement.
Longer Process Uptime, Faster Maintenance
The modular design of the InPro 6800G/12/120 allows for easy and fast membrane exchange. The sensor can be calibrated with air at the measuring point.
Improved Process Reliability
The InPro 6800G/12/120 ensures accurate, dependable measurements for a wide range of chemical and pharmaceutical applications.
Material No.: 52206425
Measurement Parameter | Gas-Phase Oxygen |
Measurement Technology | Amperometric Clark Electrode |
Measurement Range | 1 to 100% Vol.-% O2 |
Accuracy | ≤ ±[1% +0.1 Vol-%] |
Response Time | T90 ≤ 20s (nitrogen to 15 Vol-% O2 at 25 °C / 77 °F) |
Primary Media/Application | Chemical Processes Pharmaceutical/Life Science Production |
ISM (Digital) or Analog | Analog |
Cable Connector | VP7 |
Sensor Length | 120 mm |
Process Connection | Pg 13.5 |
Connector Design | Straight |
O-Ring Material | Silicone |
Sensor Diameter | 12 mm |
Surface Roughness of Wetted Parts | N5/Ra 16 (Ra=0.4mm/16min) |
Wetted Materials | 316L Stainless Steel |
Wetted membrane material | PTFE |
Model Family | InPro6800G |
Detection limit | 300 ppm |
Detection limit | 0.1 Vol.-% O3 |
Sensor signal in air at 25 °C (77 °F) | 50...110 nA |
Residual signal in oxygen free media | <0.1 % of signal in ambient air |
Temperature compensation | Automatic |
Measuring Temp. Range | 32 °F – 158 °F (0 °C – 70 °C) |
Operating Pressure (Bar) | 0.200 bar–9.000 bar |
Operating Pressure (PSI) | 2.9 to 130 psi absolute |
Mechanical Temp. Range | 23 °F – 250 °F (-5 °C – 140 °C) |
Design pressure (bar) | maximum 12 bar |
Design Pressure (PSI) | maximum 174 psi absolute |
Hazardous Area Approvals | ATEX: Ex ia IIC T6 / T5 / T4 / T3 Ga / Gb, Ex ia IIIC T69 °C / T81 °C / T109 °C / T161 °C Da / Db. FM: IS Cl. I, II, III, Div 1, GR ABCDEFG / T6 * |
Hygienic/Biocompatibility Approvals | EHEDG, FDA/USP Class VI |
Other certificates | METTLER TOLEDO Quality Certificate, Material certificate 3.1, N5/Ra 17 |
Mechanical Temperature Resistance (Celsius) | -5 °C – 121 °C |
Operating Range | 300 ppm - 100% O2 |
Connection | VP (IP 68) Pg 13.5 |
Measured Pressure Resistance | 0.2 bar – 9 bar |
Mechanical Temperature Resistance (Fahrenheit) | 23 °F – 250 °F |
Reliable in Harsh Processes
The InPro 6800G/12/120 offers accurate measurement between 0.1% and 100% oxygen, making it suitable for challenging applications such as off-gas monitoring or inertization in aggressive processes.
Suited to Your Process Needs
To meet the requirements of your specific application, the InPro 6800G/12/120 is 120 mm long for small to medium length installations. It uses a standard silicone O-ring with 316L stainless steel wetted parts.
Reduced Service Time and Fast Exchange
The InPro 6800G/12/120 uses an anode/cathode assembly with an AgCl free anode to reduce drift. The membrane body can be replaced quickly, typically in less than 2 minutes, and the sensor can be calibrated in air at the point of measurement.
Fast Response Time Ensures Safety
Amperometric oxygen sensors allow for very fast monitoring of changes in oxygen levels. The rapid response of the InPro 6800G/12/120, 90% in less than 20 seconds, allows you to quickly react to unwanted oxygen increases.