Uncover Scale-up and Safety Issues
OptiMax™ HFCal reproducibly delivers information for characterizing and scaling up processes to ensure they are safe at scale.
Understand Your Process Parameters
Heat data obtained under both isothermal and non-isothermal conditions enables the characterization of process parameters under process-like conditions.
Simple and Accurate Calorimetry
The Plug & Play calorimetry module enables determination of heat information accurately and precisely.
Material No.: 511628301
Weight | 35 kg |
Operating Volume | 150 mL – 1,000 mL |
Dimensions (HxWxD) | 78 cm x 39 cm x 42 cm |
Temperature Range | -40 °C – 180 °C |
Back / Front light | Yes/Yes |
Precision Heat Transfer | Typically +/- 3% |
Accuracy and Precision Specific Heat | Typically +/-10% |
Accuracy Heat Flow | Based on comparison of qr_hf with qc resp. ∫qr_hf with ∫ qc Isothermal conditions: ± 3 % to 5 % Non-isothermal conditions: ± 5 % to 10% |
Calorimetry Type | Heat Flow |
Material | OptiMax Starter Kit (51162830) and Upgrade Kit OptiMax HFCal (30050150) |
Temperature Modes | Jacket control, reaction mixture control, distillation control, crystallization control |
Heating / Cooling | Solid State Thermostat |
Power Requirements | 100 – 240 VAC, 50/60Hz |
Faster Process Development
Characterize reaction parameters and collect heat information at the same time. The combination of heat information with real-time analytical data provides researchers with the relevant information for safe scale-up processes.
Precise Heating and Cooling
Powerful heating and rapid cooling combined with the unique control algorithm ensure that the temperature is maintained accurately and precisely between -40°C and 180°C, even if sudden exothermic reactions occur.
Accurate Data Analysis
With a single click, the iControl software automatically calculates and reports heat transfer data, the specific heat of the reaction mass, heat flow and reaction enthalpies.
Thermostat | ||
Heating | Electrical | |
Cooling | Peltier | |
Cooling Performance | 0°C 35 W 20°C 65 W 40°C 95 W | |
Temperature Range | -40°C to 180°C (jacket temperature) | |
Control Modes | isothermal and isoperibolic, constant or ramp, reflux, distillation & crystallization | |
Stirring | 30 rpm to 1200 rpm | |
Dimensions | Thermostat: 388 mm x 414 mm x 539 mm | |
Power Supply | 100 V to 240 V AC, 50 Hz to 60 Hz, 1300 VA | |
Operation | iControl software for calorimetric applications Graphical touchscreen and/or iControl Software | |
Reactors | ||
Type | Single piece or two piece (with bottom drain valve) | |
Volume | 1000 mL | |
Pressure | 50 mbar to 1 bar | |
Material | Duran® glass | |
Stirrer | Pitch-blade (glass, Alloy 22) Anchor (glass, Alloy 22) Half-moon (PTFE for single piece reactor) | |
Calorimetry | ||
Type | Heat Flow | |
Calibration Heater | Max. 20 W | |
Precision Heat Transfer | Typically ± 3% | |
Accuracy and Precision Specific Heat | Typically ± 10% | |
Accuracy Heat Flow | Typically ± 3% to 5%, under isothermal conditions Typically ± 5% to 10%, under non-isothermal conditions | |
Sensitivity | ≤0.1 W measured as standard deviation of qr under steady state conditions. Peak-to-peak noise is ≤ approximately 0.4 W. | |
HFCal Module | ||
Power and Data Interface | Direct connection to CAN interface of standard OptiMax, no additional power supply or PC connection required | |
Data Logging | Via iControl Software | |
Dimensions | 120 mm x 40 mm x 170 mm | |
Software | iControl 5.2 or higher required, HFCal license |